RDS-30

- The RDS-80 Surface Contamination Meter is a versatile contamination detector which has been designed for a wide range of applications in different fields of radiation protection in nuclear industry, rescue and other operations, involving a possibility for abnormal contamination levels.
- The Digital, Hand Held Pulse Rate Meter is suitable for personnel who need to locate and detect contaminated surfaces/objects during their work.
- The functions comprise contamination measurements with alarm functions and automatic conversion to surface activity value and additionally surface contamination measurement with a possibility to store the histogram events into the meter’s internal memory.
- The use of configuration software (RDS-CSW ) is required for downloading the data into a PC via IrDA port.
- Item Numbers
Radiation detected: alpha >2 MeV, betaEmax >100 keV, gamma and x-rays from 5 keV to 3 MeV
Detector: end window GM tube, 1,5-2 mg/cm2
MICA window active surface area 15,2 cm2
Measurement range: surface contamination 1 to 100 000 cps or 0.01 to 1 000 000 Bq/cm2
Alarm levels: freely adjustable alarm levels for contamination level
Cps linearity: ±15% ±1 digit over the range. The Bq/cm2 display is calculated from cps values using isotope dependent coefficients
Display in either cps or Bq/cm2
Surface activity (Bq/cm2) display configurable for different isotopes
Cps rate follow-up by audible signal with frequency proportional to activity
Visual and audible alarm: user settable for surface activity
Histogram capability of up to 480 points with user settable logging interval
Backlit display with six large digits • Built-in self-diagnostics function
Gamma background reduction for Bq/cm2 measurements
Built-in infra-red port (IrDA)